Comparison of Four Mechanical Insufflation-Exsufflation Devices: Effect of Simulated Airway Collapse on Cough Peak Flow - ERPHAN - Equipe de Recherche Paramédicale sur le Handicap NeuroMoteur
Article Dans Une Revue Respiratory Care Année : 2023

Comparison of Four Mechanical Insufflation-Exsufflation Devices: Effect of Simulated Airway Collapse on Cough Peak Flow

Résumé

Background: Mechanical insufflation-exsufflation (MI-E) devices are used to improve airway clearance in individuals with acute respiratory failure. Some MI-E devices measure cough peak flow (CPF) during MI-E to optimize pressure adjustments. The aim was to compare CPF and effective cough volume (ECV: volume expired/coughed > 3 L/s) measurements between 4 MI-E devices under simulated conditions of stable versus collapsed airway. Methods: Four MI-E devices were tested on the bench. Each device was connected via a standard circuit to a collapsible tube placed in an airtight chamber that was attached to a lung model with adjustable compliance and resistance. Pressure was measured upstream and downstream the collapsing tube; air flow was measured between the chamber and the lung model. Each device was tested in 2 conditions: collapse condition (0 cm H2O) and no-collapse condition (-70 cm H2O). For each condition, 6 combinations of inspiratory/expiratory pressures were applied. CPF was measured at the "mouth level" by the device built-in flow meter and at the "tracheal level" by a dedicated pneumotachograph. Comparisons were performed with non-parametric tests. Results: CPF values measured at the tracheal level and ECV values differed between devices for each inspiratory/expiratory pressure in the collapse and no-collapse conditions (P < .001). CPF values were significantly lower at the tracheal level in the collapse as compared with the no-collapse condition (P < .001 for each device), whereas they were higher at the mouth level (P < .05) for 3 of the 4 devices. Conclusions: CPF values differed significantly across MI-E devices, highlighting limitation(s) of using only CPF values to determine cough effectiveness. In simulated of airway collapse, CPF increased at the mouth, whereas it decreased at the tracheal level.
Fichier principal
Vignette du fichier
Main_Manuscript_ R2_clean+Tab&Figs.pdf (571.89 Ko) Télécharger le fichier

Dates et versions

inserm-04064452 , version 1 (11-04-2023)

Identifiants

Citer

Nicolas Terzi, Isabelle Vaugier, Claude Guérin, Hélène Prigent, Ghilas Boussaid, et al.. Comparison of Four Mechanical Insufflation-Exsufflation Devices: Effect of Simulated Airway Collapse on Cough Peak Flow. Respiratory Care, 2023, 68 (4), pp.462-469. ⟨10.4187/respcare.10086⟩. ⟨inserm-04064452⟩
130 Consultations
102 Téléchargements

Altmetric

Partager

More